Document Type

Article

Publication Date

11-2020

Abstract

The design and construction of an instrument for full-field imaging of the X-ray fluorescence emitted by a fully illuminated sample are presented. The aim is to produce an X-ray microscope with a few micrometers spatial resolution, which does not need to scan the sample. Since the fluorescence from a spatially inhomogeneous sample may contain many fluorescence lines, the optic which will provide the magnification of the emissions must be achromatic, i.e. its optical properties must be energy-independent. The only optics which fulfill this requirement in the X-ray regime are mirrors and pinholes. The throughput of a simple pinhole is very low, so the concept of coded apertures is an attractive extension which improves the throughput by having many pinholes, and retains the achromatic property. Modified uniformly redundant arrays (MURAs) with 10 mu m openings and 50% open area have been fabricated using gold in a lithographic technique, fabricated on a 1 mu m-thick silicon nitride membrane. The gold is 25 mu m thick, offering good contrast up to 20keV. The silicon nitride is transparent down into the soft X-ray region. MURAs with various orders, from 19 up to 73, as well as their respective negative (a mask where open and closed positions are inversed compared with the original mask), have been made. Having both signs of mask will reduce near-field artifacts and make it possible to correct for any lack of contrast.

Publication Source (Journal or Book title)

JOURNAL OF SYNCHROTRON RADIATION

First Page

1703

Last Page

1706

Included in

Optics Commons

Share

COinS